RAMP Test
Jump to navigation
Jump to search
How To Run a RAMP TEST Sutter Micropipette Puller P-97
- Fo into any program. For example press 1 and ENTR to select Program 1
- Press CTRL
- Press 0 to not clear parameters
- Select 1 for RAMP TEST
- Install glass and press PULL
When executed, events in the RAMP TEST take place as follows:
- The puller increments the HEAT at the rate of 650 milliamps per second, and the filament, after 1-2 minutes, begins to get hot and glows orange in color
- The puller bars begin to move apart as the HEAT output begins to soften the glass
- The HEAT is turned off when a certain velocity (the value that is stored in ROM is achieved
- The RAMP TEST value shown in display is the HEAT value that was required to reach the factory-set RAMP TEST velocity
To interrupt the RAMP TEST or reset the display press RESET
Warning: HEAT settings over RAMP +35 can damage or burn out the filament Template:Warning