==Materials:==
+
== Material ==
<li> Ice
+
* Ice
<li> 4 mM cuvettes,  N+1 cuvettes
+
* 4 mM cuvettes,  N+1 cuvettes
<li> Steinberg's, at 4 C.
+
* Steinberg's, at 4 C.
<li> Square pulse electroporator (we use the BTX 830 Squarporator)
+
* Square pulse electroporator (we use the BTX 830 Squarporator)
  
==Procedure:==
 
# Put cuvettes on ice, and cool down Steinberg's on ice
 
# Trypsinize cells as normal, stop reaction with high serum media and spin down at 1000 rpm, 3 minutes
 
# Wash pellet with Steinberg's solution and resuspend in 300 µl of Steinberg's, keep on ice
 
# Before electroporation, make sure to electroporate a cuvette with no cells to discharge device
 
# Electroporate cells in following conditions and put back on ice: Myoblasts: 100V, 35 msec, 5 pulses, Myotubes: 75 V, 35 msec, 5pulses
 
# Plate cells as normal
 
  
 
''Steinberg's''
 
''Steinberg's''
Line 38: Line 31:
 
|Tris pH  7.8-8.0
 
|Tris pH  7.8-8.0
 
|5.75 ml (2M)
 
|5.75 ml (2M)
 +
|}
 +
 +
 +
== Procedure ==
 +
# Put cuvettes on ice, and cool down Steinberg's on ice
 +
# Trypsinize cells as normal, stop reaction with high serum media and spin down at 1000 rpm, 3 minutes
 +
# Wash pellet with Steinberg's solution and resuspend in 300 µl of Steinberg's, keep on ice
 +
# Before electroporation, make sure to electroporate a cuvette with no cells to discharge device
 +
# Electroporate cells in following conditions and put back on ice: Myoblasts: 100V, 35 msec, 5 pulses, Myotubes: 75 V, 35 msec, 5pulses
 +
# Plate cells as normal

Latest revision as of 14:25, 8 June 2015

Material

  • Ice
  • 4 mM cuvettes, N+1 cuvettes
  • Steinberg's, at 4 C.
  • Square pulse electroporator (we use the BTX 830 Squarporator)


Steinberg's

1x FW (MW) 5x/500 ml
58 mM NaCl 58.44 8.47 g
0.67 mM KCl 74.56 1.68 ml (1M)
0.44 mM Ca(NO3) 236.20 1.1 ml (1M)
1.3 mM MgSO4 246.47 3.25 ml (1M)
4.6 mM Tris pH 7.8-8.0 5.75 ml (2M)


Procedure

  1. Put cuvettes on ice, and cool down Steinberg's on ice
  2. Trypsinize cells as normal, stop reaction with high serum media and spin down at 1000 rpm, 3 minutes
  3. Wash pellet with Steinberg's solution and resuspend in 300 µl of Steinberg's, keep on ice
  4. Before electroporation, make sure to electroporate a cuvette with no cells to discharge device
  5. Electroporate cells in following conditions and put back on ice: Myoblasts: 100V, 35 msec, 5 pulses, Myotubes: 75 V, 35 msec, 5pulses
  6. Plate cells as normal